USER'S MANUAL AE-966 / 967
6.2.2

Amplitude

Amplitude key sets the vertical scale of the display, including the upper limit
(Reference Level), vertical range/unit (Amplitude Scale and Unit), and compensation for
external gain or loss (External Offset). Amplitude Correction adjusts the frequency
response distortion caused by external networks. Pre-Amplifier is an optional item that
boosts the level of weak input signal before entering GSP-830. You can also set input
impedance level according to the application needs.
6.2.2.1
Set Vertical Scale.
Vertical display scale is defined by the reference amplitude level, vertical
amplitude range, measurement unit, and external gain/loss setting.
Set reference amplitude level
The reference level defines the uppermost display level.
Panel
1.- Press the Amplitude key.
operation
2.- Press F2 (Start).
3.- Enter the value using the numerical keys,
Arrow key, and Scroll knob.
Range
dBm -110 ~ +20 dBm, 0.1dB resolution.
dBmV -63.01 ~ +66.99 dBmV, 0.01dB resolution.
dBμV -3.01 ~ +126.99 dBuV, 0.01dB resolution.
04/2016
Amplitude
Ref.Level
F 1
-30.0dBm
μ
Page 39