Example: after
Start frequency: 97.5MHz
Autoset
Span: 5MHz
Center frequency: 100MHz
6.2.3.2
RBW/VBW/Sweep Setting after Autoset
All the three BW related parameters, RBW, VBW, and Sweep, will be reset to
Auto mode when using Autoset, regardless of their previous settings
•
Limit vertical search range
Background You can set the amplitude floor so that the signals lower than the
setting will be ignored from Autoset search.
Panel
1.- Press the Autoset key.
operation
2.- Press F2 (Amp. Floor) to switch the range
from Auto (whole range) to manual
(limited range).
3.- Enter the amplitude in dB, using the
numerical keys. The Command window
shows the setting.
Range
dBm -80 ~ +20dBm, 0.1dB resolution.
dBmV -33.01 ~ +66.99dBmV, 0.01dB resolution.
dBμV +26.99 ~ +126.99dBμV, 0.01dB resolution.
Page 52
USER'S MANUAL AE-966 / 967
Stop frequency: 102.5MHz
Signal peak: 100MHz
Reference level: -30dBm
Autoset
Amp.Floor
Auto Man
μ
Amp. : -30dBm
F 2
04/2016