Promax AE-966 Manual De Instrucciones página 361

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

meas:lmtline:edit
meas:lmtline:edit:
delall
BW
Comand
con:rbw:auto
con:rbw?
con:rbw:man
con:rbw:mode?
con:vbw:auto
con:vbw?
con:vbw:man
con:vbw:mode?
con:swt:auto
con:swt:man
Page 170
Sets the limit line table data. Need to specify 0 (low limit
line), 1(high limit line) / limit line points.
Example:
meas:lmtline:edit_0_CR_3,100, 20,110, 30,120, 25 (low
limit line, 3 points, 100MHz/ 20dB, 110MHz/ 30dB,
120MHz/ 25dB)
Delete all points in limit line table.
Parameter: 0 (low limit line), 1(high limit line)
Example: meas:lmtline:edit:delall_0 (delete low lline table)
Sets RBW to auto.
Returns the RBW (resolution bandwidth).
Parameter: 0 (10kHz), 1 (300Hz), 2 (3kHz), 3 (9kHz), 4
(30kHz), 5 (120kHz), 6 (300kHz), 7 (4MHz), 8 (100kHz)
Selects the RBW.
Parameter: 0 (200kHz), 1 (300kHz), 2 (3kHz), 3 (9kHz), 4
(30kHz), 5 (120kHz), 6 (300kHz), 7 (4MHz)
Example: con:rbw:man_1 (sets RBW to 300Hz)
Returns RBW mode. Parameter: auto, manual
Sets VBW to auto.
Returns the VBW (video bandwidth).
Parameter: 0 (100Hz), 1 (1kHz), 2 (3kHz), 3 (10kHz), 4
(30kHz), 5 (100kHz), 6 (300kHz), 7 (4MHz)
Selects the VBW.
Parameter: 0 (100Hz), 1 (1kHz), 2 (3kHz), 3 (10kHz), 4
(30kHz), 5 (100kHz), 6 (300kHz), 7 (4MHz)
Example: con:vbw:man_1 (sets VBW to 1kHz)
Returns VBW mode.
Parameter: auto, manual
Sets the sweep time to auto.
Sets the sweep time in msec.
Example: con:swp:man_5 (sets the sweep time to 5ms)
USER'S MANUAL AE-966 / 967
Description
04/2016

Publicidad