Select 9Khz/120Khz Rbw; Check Emi Filter Installation Status - Promax AE-966 Manual De Instrucciones

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

USER'S MANUAL AE-966 / 967

6.2.18.2 Select 9kHz/120kHz RBW

Panel
1.- Press the BW key.
operation
2.- Press F1 (RBW Manu).
3.- Select 9kHz/120kHz RBW using the Scroll
key. The RBW value appears in the command
window.
Note
In the automatic mode, RBW is selected according to the internal
reference setting.

6.2.18.3 Check EMI filter installation status

Panel
1.- Press the System key..
operation
2.- Press F6 (More).
3.- Press F4 (System Config On). The system
configuration window appears.
4.- The 9/120k RBW sign shows the installation
status, installed (checked) or not installed
(unchecked).
Note
EMI filter (9k/120k RBW) and 10k/100k RBW are exclusive. They cannot
be installed together.
04/2016
BW
RBW
F 1
Auto Manu
RBW: 9kHz
System
More...
F 6
System Config
F 4
On
Off
Installed
Not installed
Page 143

Publicidad

Capítulos

Tabla de contenido
loading

Este manual también es adecuado para:

Ae-967

Tabla de contenido