H. CLEANING, DISINFECTION, AND STERILIZATION
For hygiene and sanitary safety purposes, the contra-angle attachments must be
cleaned, disinfected, and sterilized before each usage to prevent any contamination.
This concerns the first use, as well as all subsequent uses.
PRECAUTION
Do not sterilize (autoclave) the motor handpiece, charging stand, and AC cable. The handpiece
body is not waterproof. Do not immerse the handpiece body, charging stand or AC cable in
water, or a chemical solution. Do not spray liquids onto this device – wipe only.
GENERAL RECOMMENDATIONS
1. Use only a disinfecting solution which is approved for its efficacy (VAH/
DGHM-listing, CE marking, FDA and Health Canada approval) and in
accordance with the DFU of the disinfecting solution manufacturer.
a. Do not use chloride detergent materials.
b. Do not use bleach or chloride disinfectant materials.
2. For your own safety, please wear personal protective equipment (gloves,
glasses, mask).
3. The user is responsible for the sterility of the product for the first cycle and each
further usage as well as for the usage of damaged or dirty instruments where
applicable after sterility.
4. The water quality has to be convenient to the local regulations especially for
the last rinsing step or with a washer-disinfector.
5. Do not sterilize the motor handpiece, the charging stand, or the AC cable. After
each use, all the objects that were in contact with infectious agents should be
cleaned using towels impregnated with a disinfecting and detergent solution
(a bactericidal, fungicidal, and aldehyde free solution) approved by either
VAH/DGHM-listing, CE marking, FDA, or Health Canada.
19