Before using and/or after cleaning, pay special attention to ensure that all the accessories supplied with the device are intact.
Switch off the device before cleaning it and disconnect the mains cable from the electrical socket.
PREPARATION
1.
Pull out the air tube from the nebulizer and leave it plugged into the air outlet nozzle of the device;
2.
Rotate the upper part of the nebulizer anti-clockwise;
3.
Use your fingers to disconnect the internal pisper at the bottom of the nebuliser.
CLEANING
Before and after each use proceed with cleaning all of the components of the nebulizer (with the exception of the air tube)
according to one of the two methods described below.
Method 1: Thoroughly clean the components for 5 minutes, using warm drinking tap water (about 40°C) and/or mild soap.
Method 2: Clean the components (except for the air tube) by immersing them in a solution with 60% water and 40% white
vinegar. When finished, thoroughly rinse with warm drinking water (approx. 40°C).
After cleaning, rinse thoroughly by removing the excess water and allow to air dry in a clean place.
DO NOT BOIL OR AUTOCLAVE THE AIR TUBE AND MASKS
DO NOT WASH ACCESSORIES IN A DISHWASHER
WASHING
If there are pathologies with risks of infection and microbial contamination, it is the end user's responsibility to proceed with
suitable washing. The washing procedure can only be carried out if the components to be treated have undergone specific cleaning
(see chapter on cleaning).
Proceed as follows for the washing procedure:
- Fill a container, of a suitable size to contain all the individual components, with a solution of drinking water and disinfectant
(hypochlorite-based solution readily available in a pharmacy) by following the proportions indicated on the packaging of the
disinfectant itself;
- The period of time for which is to be immersed in this solution is indicated on the packaging of the hypochlorite solution in
accordance with the chosen concentration for preparing the solution;
- Rinse thoroughly with lukewarm drinking water to remove all traces of the solution.
Dry and store in a dry, dust-free environment.
- Dispose of the used solution according to the instructions provided by the manufacturer of the disinfectant solution.
For each individual patient it's recommended to use the nebulizer for 6 months or for a maximum of 120 treatments.
The nebulizer must be replaced after a long period of inactivity, if it is deformed or broken, or if the nebulizer nozzle is blocked by
dry medicine, dust, ecc.. Only use the original nebulizer supplied by GIMA with the device
Use the "nose piece" accessory only if expressly indicated by your doctor and paying attention NEVER to introduce inside the nose
the nasal bifurcation, but only bring it as close as possible.
In the presence of infection or microbial contamination prone pathologies, we recommend using your personal accessories
and nebulizer (always consult your doctor).
must be replaced every 25 hours of functioning or when it result particularly worn. Only use original GIMA filter.
The device is equipped with a filter that removes any impurities from the air that was sucked in by the compressor. The air filter
The mask and tube must be replaced as soon as the materials they are made of show signs of deterioration.
DON'T USE THE DEVICE WITHOUT AIR FILTER
CLEANING AND WASHING OF ACCESSORIES
STANDARD ACCESSORIES
HI-FLO KIT – REF RE 300300
(Nebulizer HI-FLO, Adult Mask, Pediatric Mask, Air Tube and Mouth-piece,
ACCESSORIES
Nosepiece)
1-
2-
3-
4-
5-
6-
7-
8-
13
Air Tube
Nebulizer Tank
Nebulization Nozzle
Nebulizer Top
Mouthpiece
Adult Mask
Pediatric Mask
Nosepiece
GB