Table 6-2. Instrument Specifications
OPERATING
-20 to 50°C (-4 to +122°F), [NH 3 , ClO 2 : -20 to 40°C (-4 to 104°F),
TEMPERATURE
O 2 -R Version: 0 to 40°C (32 to 104°F)]
RANGE
HUMIDITY
10 to 95% RH, Non condensing
INGRESS
IP67, (Remote O 2 is IP54)
PROTECTION
RECOMMENDED 0 to 40°C (32° to 104°F)
STORAGE
WARRANTY
See Chapter 5
AUDIBLE
95 dB typical
ALARM
APPROXIMATE
3.4" H x 2.0" W x 1" D
SIZE
(8.6 cm H x 5.1 cm W x 2.5 cm D)
WEIGHT
4 oz (113 grams)
SENSOR
Electrochemical Sensors
BATTERY
Lithium, CR2, non-rechargeable:
Replace with Energizer CR2 EL1CR2BP, Panasonic CR-2PA/1B,
Varta CR2 6206, or Varta (Power One) CR2 CR15H270, 6206
battery only
FACTORY-SET*
ALARM
SETPOINTS
LOW*
HIGH*
ALARM ALARM (PPM)
(PPM)
(PPM)
CO
& CO
FIRE
25
100
CO
STEEL 75
200
H 2 S
10
15
O 2
19.5%
23.0%
SO 2
2.0
5.0
NO 2
2.0
5.0
NH 3
25
50
PH 3
0.3
1.0
HCN
4.5
10.0
Cl 2
0.5
1.0
ClO 2
0.10
0.30
NOTE:
This instrument has not been classified for use in atmospheres containing >21% oxygen.
* Other setpoints available upon request or at any time via MSA FiveStar Link Software.
6-2
STEL*
TWA* MIN.
(PPM) ALARM ALARM
SET-
POINT
100
25
20
200
75
20
15
10
5
--
--
5.0
5.0
2.0
2.0
5.0
2.0
2.0
35
25
15
1.0
0.3
0.3
10.0
4.5
4.5
1.0
0.5
0.5
0.30
0.10
0.10
MAX.
FAS
CAL/
TIME
SPAN
SET-
(SEC)
TIME
POINT
(SEC)
1450
<15
90
1450
<15
90
175
<15
90
24
<15
90
17.5
<15
90
17.5
<15
240
75
<15
180
3.75
<15
90
20.0
<15
240
17.5
<15
240
0.75
<15
360