5. Once the instrument successfully calibrates and the "OK" displays:
"CAL" "?" displays.
6. While the "CAL" "?" displays, press the TEST button to enter the
Gas Calibration mode.
The current expected test gas is shown (in ppm).
a. To change the expected calibration gas:
1) Press the TEST button.
"TEST" "SET" "?" "ppm" displays.
2) Press the TEST button in rapid succession to increment.
The TEST button can be held down to increment by
larger values..
3) Wait three seconds to return to the Calibration mode.
b. Apply the appropriate test gas
The display toggles between the current gas reading, the
hourglass and "CAL".
Once the instrument passes calibration, "OK" displays.
Calibration time is dependent on the sensor installed. See
TABLE 6-2.
Otherwise, "ERR" displays.
Wait five seconds to return to the normal operating mode.
The expected gas concentrations must match the gas con-
centrations listed on the calibration cylinder(s). Failure to fol-
low this warning will cause an incorrect calibration, which
can result in serious personal injury or death.
c. If "ERR" displays after calibration, the current settings did not
change. Immediately check that:
The calibration cylinder matches the expected calibration
value expected in the instrument
The calibration cylinder is not empty and has not expired.
The regulator is 0.25 liters/minute
The tubing is seated in the front instrument case during
gas calibration mode.
If necessary, repeat steps 1 through 6.
The display must read "OK"; if "ERR" remains, remove the
instrument from service.
d. Perform a bump test to confirm operation and activate the "√".

" WARNING

4-3