SENSOR:
RANGE:
RESOLUTION:
SENSOR:
RANGE:
RESOLUTION:
* Normal temperature range 0-40°C
** extended temp. range -20 to 0, 40 to 50°C (NH 3 and ClO 2 only: -20 to 0°C)
Table 6-5. Data Logging Specifications
SESSION
NUMBER OF
DATALOG
SHARED EVENTS
DATA TRANSMISSION
METHOD
EVENT LOG
INFORMATION
TRANSMISSION TIME
PERIODIC
DEFAULT LOGGING
DATALOG
INTERVAL
STORAGE ESTIMATE
TRANSMISSION TIME
6-4
Cl 2 (Chlorine)
0-20.0 ppm
0.1 ppm
ClO 2 (Chlorine Dioxide)
0-1.00 ppm
0.02 ppm
50 (most recent occurrences)
Via MSA infrared adapter on a PC
using MSA FiveStar Link Software >
version 4.5
Alarm - Alarm Type - Alarm Value -
Time/Date
Alarm Clear - Alarm Type - Alarm Value
- Time/Date
Cal (Pass/Fail) - Time/Date
Bump (Pass/Fail) - Time/Date
Error Non-Shutdown - Error Type
(See Error List) - Time/Date
ON/OFF - Time/Date.
Typically less than 60 seconds max.
Three minute peak readings
(configurable via PC from 15 seconds
to 15 minute peaks or one minute peak
average)
Typically greater than 100 hours with
default interval (storage time varies
based on interval and sensor activity)
Typically less than three minutes