Data Logging; Session Log; Periodic Log - MSA ALTAIR Pro Manual De Operación

Ocultar thumbs Ver también para ALTAIR Pro:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Data Logging

Session Log

The ALTAIR Pro Single Gas instrument has the capability of
recording 50 of the most recent events.
To transfer data to a pc, hold the TEST button for one second.
The instrument pages begin to display.
Put the instrument in IR mode (see FIGURE 2-5).
Events will be transferred to the PC during this sequence, if:
the top of the instrument is pointed toward the optional IR
Receiving device
a PC is running MSA FiveStar Link Software (P/N 710946)
"CONNECT" is pressed in the PC FiveStar LINK Software
package.
The following events are recorded:
Alarm - Alarm Type - Alarm Value - Time/Date
Alarm Clear - Alarm Type - Alarm Value - Time/Date
Cal (Pass/Fail) - Time/Date
Bump (Pass/Fail) - Time/Date
Error - Error Type (See Error List) - Time/Date

Periodic Log

Peak gas readings are periodically logged, based on user settings
via the PC. (Default is one sample every three minutes.)
See TABLE 6-3 for typical performance.
NOTES:
The time and date is based on the PC time and date.
Ensure the PC is correct before communicating with the
instrument.
Changing the instrument battery can result in lost time in
the Data log.
After changing batteries, verify the Time and Date with
your PC.
2-17

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Altair pro

Tabla de contenido