Immunity test
Power
frequency
(50/60 Hz)
magnetic field
(IEC
61000-4-8)
a. The LATITUDE Communicator is intended for use in the
electromagnetic environment specified above. The customer or
the user of the LATITUDE Communicator should assure that it is
used in such an environment.
b. U
is the AC main's voltage prior to application of the test level.
T
Table 3. Guidance and manufacturer's declaration—
electromagnetic immunity—for equipment and systems that are
not life supporting
IEC 60601
Immunity test
test level
90
LATITUDE
IEC 60601 test
Compliance level
level
3 A/m
30 A/m
a b c d
Compliance
level
Communicator Patient Manual
TM
Electromagnetic
environment—guidance
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial
or hospital
environment.
Electromagnetic environment—
guidance
Portable and mobile
RF communications
equipment should be used
no closer to any part of the
LATITUDE Communicator,
including cables, than the
recommended separation
distance calculated from
the equation applicable
to the frequency of the
transmitter.
Recommended separation
distances: