IEC 60601
Immunity test
test level
Radiated
3 V/m
RF (IEC
80 MHz to
61000-4-3)
2.5 GHz
a.
The LATITUDE Communicator is intended for use in the electromagnetic
environment specified above. The customer or the user of the LATITUDE
Communicator should assure that it is used in such an environment.
b.
At 80 MHz and 1 GHz, the higher frequency range applies.
92
LATITUDE
Compliance
level
10 V/m
80 MHz to
1000 MHz
3 V/m
1 GHz to
2.7 GHz
8 V/m
5.1 GHz to
5.8 GHz
Communicator Patient Manual
TM
Electromagnetic environment—
guidance
d = 0.5 √P (80 MHz to 1000
MHz)
d = 0.29 √P (380 MHz to 390
h
MHz)
d = 0.36 √P (430 MHz to 470
MHz)
d = 0.21 √P (800 MHz to 960
i
MHz)
d = 1.7 √P (1 GHz to 2.7 GHz)
d = 0.21 √P (1.7 GHz to 1.99
GHz)
d = 0.21 √P (2.4 GHz to 2.57
GHz)
d = 0.63 √P (5.1 GHz to 5.8 GHz)
where P is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and d is the recommended
separation distance in
meters (m).
Field strengths from
fixed RF transmitters,
as determined by an
electromagnetic site
survey,
should be less
f
than the compliance level
in each frequency range.
Interference may occur in
the vicinity of equipment
marked with the following
symbol:
Non-iodizing
electromagnetic
radiation
g