2
Technical Data and Further Information
The most important technical data such as supply voltage are given on the name plate either
outside or inside the transmitter housing. For further technical data such as accuracy refer to
the Operation Manual. This document, the Operation Manual and software are on the supplied
CD-ROM. You can also download the documentation via Internet "www.mt.com/M400".
3
Mounting
The M400 transmitter is available is available as 1/2 DIN version.
For installation drawings refer to the Operation Manual.
1. Mount the supplied cable glands at the housing.
2. Mount the transmitter. You have the following possibilities:
• Panel mounting, wall mounting or pipe mounting.
4
Electrical Connection
h
Note: Power off device during electrical connection.
1. Switch off supply voltage.
2. Connect mains supply to the terminals +FF-H1 and –FF-H1.
• FISCO bus: 9 to 17.5 V DC
• Linear barrier: 9 to 24 V DC or
• Non hazardous area (Non-IS) : 9 to 32 V DC
3. Connect sensor to terminal block TB2 according to the Operation Manual..
en
35