Detailed Information About Your System - OneTouch VerioIQ Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

8
Detailed information
about your system
Comparing meter results to laboratory
results
Results obtained from the OneTouch Verio IQ® Meter and
laboratory tests are reported in plasma-equivalent units.
However, your meter result may differ from your lab result
due to normal variation. A result from your OneTouch
Verio IQ® Meter is considered accurate when it is within
±15% of the lab result.
Meter results can be affected by factors that do not affect
lab results in the same way, which may cause a difference
of more than ±15%. Specific factors that may cause your
meter result to vary from your lab result by more than
±15% include:
• You have eaten recently. This can cause a result from
fingertip testing to be up to 70 mg/dL higher than a lab
test using blood drawn from a vein.
• Your hematocrit is above 60% or below 20%.
• You are severely dehydrated.
• For additional information, refer to the OneTouch Verio®
Test Strip Insert.
1. Sacks, D.B.: "Carbohydrates." Burtis, C.A., and Ashwood E.R. (ed.),
Tietz Textbook of Clinical Chemistry, Philadelphia: W.B. Saunders
Company (1994), 959.
1
102

Publicidad

Capítulos

Tabla de contenido
loading

Tabla de contenido