Detailed information about your system
System Accuracy
Diabetes experts have suggested that glucose meters
should agree within 15 mg/dL of a laboratory method
when the glucose concentration is lower than 75 mg/dL,
and within 20% of a laboratory method when the glucose
concentration is 75 mg/dL or higher. Samples from
100 patients were tested using both the OneTouch
Verio IQ® System and the YSI 2300 Glucose Analyzer
laboratory instrument.
System Accuracy Results for
Glucose Concentrations <75 mg/dL
Percent (and number) of meter results that match the
laboratory test
Within
±5 mg/dL
88.2%
(45/51)
Within
±10 mg/dL
100%
(51/51)
106
Within
±15 mg/dL
100%
(51/51)