Calibrate The Analyzer; Operation; Configuration - Hach APA 6000 Manual De Operación Básica

Ocultar thumbs Ver también para APA 6000:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

1. Push MENU and select SENSOR MENU>APA6000 #1 APA
6000>SOLUTIONS AND CLEANING>PRIMING>PRIME ENTIRE
SYSTEM NOW to prime the analyzer.
Priming takes 7–15 minutes. A pop-up window shows the time that
remains to complete the prime cycle. When the priming is complete,
the analyzer automatically starts sample measurement.
2. If bubbles are seen in the tubing after the analyzer is primed, do step
1 again to remove the bubbles (maximum of 4 times).

Calibrate the analyzer

Calibrate the analyzer before initial use. The manufacturer recommends
that the analyzer be calibrated after at least 5 hours of operation to let all
system components stabilize. Refer to
on page 11 to start a calibration.

Operation

Configuration

Set the language, date, time and analyzer name
1. Push MENU and select AQUATREND MENU>LANGUAGE.
2. Select the language.
3. Select SET TIME/DATE.
4. Select an option.
Option
Description
SET TIME
Enter the time in 24-hour format (HH:MM).
SET DATE
Enter the date.
DAY
Enter the day of the week.
DATE FORMAT
Select the date format (MM:DD:YY or DD:MM:YY).
5. Optional: Change the default name for the analyzer (default =
#1 APA 6000).
a. Select EDIT NAME.
Start a calibration manually
b. Use the arrow keys to change the name, then push ENTER. The
name can be a maximum of 12 alphanumeric characters. The
first character must be a letter or #.
Configure the display settings
1. Push MENU and select AQUATREND MENU>DISPLAY SETUP.
2. Select an option.
Option
Description
MEASUREMENT TO
Assign a channel to each parameter measured
CHANNEL>ATTACH
on a sample stream—CHANNEL A–P. Only
parameters assigned to channels show on
the display.
Note: To change an assigned channel, detach
the channel first and then attach the channel to
a different parameter.
CHANNEL GRAPH
Set the minimum (YMIN) and maximum (YMAX)
SETTINGS
y-axis graph values for each channel. To enter
a negative value, change the first character to a
negative sign.
FLASH ON ALARMS
Enable or disable the display to flash when
there is an active alarm—YES (enable) or NO
(disable).
CONTRAST LEVEL
Set the display contrast—00 to 24 (default =
14).
Select the measurement units
Select the measurement units for the parameters shown on the display.
Note: The data log for a parameter is erased when the measurement units for the
parameter are changed.
English 7

Publicidad

Tabla de contenido
loading

Tabla de contenido