3.2.2 Description of the remote control commands

758 KFD Titrino
3.2.2.1.
Mode
Start and stop ($G, $S) or hold of the current method (3.2.2.3) with $H and
continue with $C.
$G also serves to continue after inquiries of identifications and sample size af-
ter the start (see 3.2.2.28) as well as after inquiries of calibration temperature
and pH values of buffers (see 3.2.2.39 and 3.2.2.40).
3.2.2.2.
Mode.QuickMeas
Start and stop of a measurement in the basic mode with the parameters
(measured quantity, measuring input) of the current method. Corresponds to
the <meas/hold> key. In TIP the measured quantity is selected with
&Mode.Parameter, see 3.2.2.47.
With an ongoing measurement, the current mode can be started. This stops
the measurement automatically.
3.2.2.3.
Mode.Select
Mode.SETQuantity
Mode.MEASQuantity
Mode.KFTQuantity
Selection of the standard mode. Mode and the measured quantity belong to
the complete selection.
If a method is selected from the method memory, the nodes &Mode.Select
and &Mode.XXXQuantity are overwritten with mode and measured quantity of
the corresponding user method.
3.2.2.4.
Mode.Name
Name of the current method in the working memory. $Q sends 8 ASCII char-
acters. Standard methods carry the name ********. The node can be set
read + write, see 3.2.2.105.
3.2.2.5.
Mode.Def.Formulas.1.Formula
Mode.Def.Formulas.1.TextRS
Mode.Def.Formulas.1.Decimal
Mode.Def.Formulas.1.Unit
Mode.Def.Formulas.1.Limits
Mode.Def.Formulas.1.LoLim
Mode.Def.Formulas.1.UpLim
Mode.Def.Formulas.1.Output
Mode.Def.Formulas.2.Formula
etc. up to .9
Entry of formulas. Rules for formula entry, see page 42.
Example: "(EP2–EP1)∗C01/C00"
In addition to the formula, a text for result output, the number of decimal
places and a unit for the result output can be selected. "No unit" is selected
with the blank string.
3.2 Remote control commands
$G, $S, $H, $C
KFT, SET, MEAS, CAL, TIP
pH, U, Ipol, Upol
pH, U, Ipol, Upol, T
EPX, CXX, RSX, +, –, ∗, /, (, )
up to 8 ASCII characters
up to 6 ASCII characters
0...±999 999
0...±999 999
active, pulse, OFF
$G, $S
Ipol, Upol
read only
0...2...5
ON, OFF
99