758 KFD Titrino
3.2.2.113. Info.ActualInfo.Titrator.CyclNo
Info.ActualInfo.Titrator.V
Info.ActualInfo.Titrator.Meas
Info.ActualInfo.Titrator.dVdt
Info.ActualInfo.Titrator.dMeasdt
Info.ActualInfo.Titrator.dMeasdV
Info.ActualInfo.Titrator.T
$Q sends the current values in the following formats:
CyclNo
V(mL)
Meas:
pH
U, Ipol (mV)
Upol (µA)
T (°C)
dVdt (µL/s)
dMeasdt
pH,U,Ipol (mV/s)
Upol (µA/s)
T (°C/s)
dMeasdV (mV/µL)
T (°C)
NV: Not Valid. If in the signal drift is OFF in modes MEAS and CAL, the signal
drift is NV.
OV will be sent for "overrange".
A time frame can be set up from the cycle number and the cycle time (see
3.2.2.118). The cycle number is set to 0 at the start of a method and it is in-
cremented until the end of the method.
3.2.2.114. Info.ActualInfo.MeasPt.Index
Info.ActualInfo.MeasPt.X
Info.ActualInfo.MeasPt.Y
Info.ActualInfo.MeasPt.Z1
Info.ActualInfo.MeasPt.Z2
Info.ActualInfo.EP.Index
Info.ActualInfo.EP.X
Info.ActualInfo.EP.Y
$Q sends the last entry into the measuring point list (.MeasPt) or the last entry
into the list of EP's with SET, KFT.
.MeasPt.X"165" Time (SET, KFT, MEAS) of the MPList in s
.MeasPt.Y"3.654" Volume (SET, KFT) of the MPList in mL
.MeasPt.Z1"6.34" Measured value (SET, KFT, MEAS) of the MPList, format
depends on the measured quantity
.MeasPt.Z2"25.8" Temperature of the MPList in °C
.EP.X"1.234" Volume coordinate of the EP
.EP.Y"5.34" Measured value coordinate of the EP
3.2 Remote control commands
SET
KFT
127
127
1.2345
1.2345
3.6(mV)
-
–345.6
–345.6
–12.5
–12.5
-
-
2.5142
2.5142
0.7957
0.7957
0.7957
0.7957
-
-
10.6326
10.6326
-
-
read only
read only
read only
read only
read only
read only
read only
MEAS
CAL
127
127
-
-
3.345
3.345
–345.6
-
–12.5
-
25.0
-
-
-
0.7957
0.7957
0.7957
-
0.7957
-
-
-
25.9
-
read only
read only
read only
read only
read only
read only
read only
read only
121