3.2 Remote control commands
124
3.2.2.126. Assembly.Outputs.AutoEOD
Assembly.Outputs.SetLines
Assembly.Outputs.SetLines.L0
up to .L13
Assembly.Outputs.ResetLines
Setting the I/O output lines.
.AutoEOD: The automatic output of the EOD (End of Determination) at the
end of the determination can be switched off. Thus, for example,
in conjunction with a Titrino several determinations can be per-
formed in the same beaker. Before AutoEOD is switched on, line 3
must be set to "OFF".
.SetLines: With $G, all lines are set.
.SetLines.LX: Set the line LX. "active" means setting of a static signal, "inac-
tive" means resetting of the signal, "pulse" means output of a
pulse of app. 150 ms, "OFF" means the line is not operated, see
also page 174.
Warnings:
• If you have "AutoEOD" to "ON", an active line 3 is set to "inac-
tive" by the EOD pulse.
• L6 is the line of the activate pulse. An active line 6 is set to "in-
active" by the activate pulse.
• L5 is the error line. It is continuously controlled by the Titrino
program and can therefore not be set freely.
Line assignments in Titrino program:
L0
Ready, inactive state
L1
Conditioning OK
L2
Titration in progress
L3
EOD (End Of Determination)
L4
Can be set in TIP
L5
Error
L6
Activate pulse + can be set in TIP
L7
Buret volume pulses
L8–13
---
.ResetLines: Lines are set to the inactive status (= high).
3.2.2.127. Assembly.Stirrer
Switching stirrer ON/OFF.
3.2.2.128. Setup.Comport
Selects the Titrino COM for the output of automatic info:
&Setup.Keycode
&Setup.Trace
&Setup.SendMeas
&Setup.AutoInfo
ON, OFF
active, inactive, pulse, OFF
ON, OFF
1, 2, 1&2
758 KFD Titrino
$G
$G