3.2 Remote control commands
118
Recalculation of the measuring data is triggered with $G.
.Write:
With "ON", the following nodes can be overwritten:
&Info.DetermData.MP.List, &Info.TitrResults.Var.C4X (X = 0...5),
&Info.TitrResults.TempVar.C7X (X = 0...9), and &Mode.Name.
.ExV:
Volume of the exchange unit, with which the determination was
executed
.MPList.1.Attribute: Attribute
.MPList.X: X coordinate, time
.MPList.Y: Y coordinate, volume
.MPList.Z1: Z1 coordinate, measuring value
.MPList.Z2: Z2 coordinate, temperature
3.2.2.106. Info.TitrResults.RS.1.Value
etc., up to .9
Info.TitrResults.EP.1.V
Info.TitrResults.EP.1.Meas
etc., up to .2
Info.TitrResults.Var.C40
etc., up to .C47
Info.TitrResults.Var.DTime
Info.TitrResults.TempVar.C70
etc. up to .C79
.RS: Values of the calculated results.
.EP: Endpoints with SET, KFT:
Volume coordinate in mL, e.g. "1.2340"
Measured value coordinate in pH "5.12", mV (with U and Ipol)
"–241" or µA (with Upol) "43.7".
.Var: Various variables. You may overwrite the variables C40...C45, see
3.2.2.105.
C40: Initial measured value in pH "5.12", mV (with U and Ipol)
"41", µA (with Upol) "43.7" or °C (with T) "25.0". In MEAS final
measured value.
C41: End volume with SET and KFT in mL, "12.5360".
C42: Time from start of titration to end in s, "62".
C43: Volume drift on start of a SET/KFT titration from the condi-
tioning in µL/min, "3.5".
C44: (Last measured) temperature in °C. Used for the temperature
compensation in pH measurements.
C45: Start volume with SET and KFT in mL, "2.800".
C46: Asymmetry pH of CAL, "6.89".
C47: Relative electrode slope of CAL, "0.9950".
C48: Voltage at maximum voltage (not valid in CAL and TIP)
C49: Voltage at minimum voltage (not valid in CAL and TIP)
DTime: Time for the drift correction in SET/KFT with conditioning.
.TempVar: Temporary variables in TIP corresponding to the assignments in
the submethods.
read only
read only
read only
read only/read+write
read only/read+write
read only/read+write
758 KFD Titrino