758 KFD Titrino
.Right:
Relative specification of the width of the output medium (e.g. pa-
per width) for the length of the measured value axis. 1 means the
measured value axis is plotted over the entire width of the paper
(largest possible width). In extreme cases, the writing of the right
tick may lie outside.
.Feed:
Length of the time axis. Depending on the printer, the measure in
cm may not always be correct.
Curve length
0.01
100 cm
0.1
0.5
1
3.2.2.140. Setup.PowerOn
Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.
3.2.2.141. Setup.Initialise
Setup.Initialise.Select
Setting of default values for the following areas:
ActMeth:
Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Silo:
The silo memory is deleted. Same function as delete entire silo.
Calib:
pH calibration data for all measuring inputs.
Config:
All values under &Config.
Assembly: All values under &Assembly.
Setup:
All values under &Setup.
All:
Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
3.2.2.142. Setup.RamInit
Initializes instrument, see page 153. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted. The user memory contains the default user methods from Metrohm.
3.2.2.143. Setup.InstrNo
Setup.InstrNo.Value
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
3.2.2.144. Diagnose.Report
Output of the report containing the adjustment parameters. The Titrino has to
be in its inactive basic state.
3.2 Remote control commands
10 cm
2 cm
1 cm
serial number, 8 ASCII characters
ActMeth, Silo, Calib, Config,
Assembly, Setup, All
$G
$G
$G
$G
$G
129