3.2 Remote control commands
110
3.2.2.60. MemoryCard.List.Card.1.Name
MemoryCard.List.Card.1.Bytes
for each directory
List of all directories on the memory card with the following information:
.Name:
Name of the directory
.Bytes:
Number of bytes used by the directory
3.2.2.61. MemoryCard.List.ActDir.1.Name
MemoryCard.List.ActDir.1.Mode
MemoryCard.List.ActDir.1.Quantity
MemoryCard.List.ActDir.1.DosUnit
MemoryCard.List.ActDir.1.Bytes
MemoryCard.List.ActDir.1.Checksum
for each method
List of all methods of the current card directory with the following information:
.Name:
Name of the method
.Mode:
Mode
.Quantity: Measured quantity
.DosUnit: Buret of the method
.Bytes:
Number of bytes used by the method
.Checksum: Checksum of the method, see 3.2.2.104.
3.2.2.62. Config.Monitoring.Validation.Status
Config.Monitoring.Validation.Interval
Config.Monitoring.Validation.Counter
Config.Monitoring.Validation.ClearCount
Monitoring of validation.
.Interval:
Time interval in days for validation.
.Counter:
Time counter in days since last validation.
.ClearCount: Clears the above counter.
3.2.2.63. Config.Monitoring.Calibration.Status
Config.Monitoring.Calibration.MeasInput
Config.Monitoring.Calibration.Interval
Config.Monitoring.Calibration.Counter
Monitoring of pH calibration.
.MeasInput:Measuring input.
.Interval:
Time interval in days for calibration from the last calibration date.
.Counter:
Time counter in days since last calibration. The counter is reset to
zero if a new calibration is carried out or if the calibration for the
corresponding measuring input is entered manually.
3.2.2.64. Config.Monitoring.Service.Status
Config.Monitoring.Service.Date
Monitoring of service interval.
read only
read only
read only
read only
read only
read only
read only
read only
ON, OFF
1...365...9999
0...9999
$G
ON, OFF
1, 2, diff
1...7...9999
0...9999
ON, OFF
YYYY-MM-DD