Pin Assignment - Metrohm 758 KFD Titrino Instrucciones Para El Uso

Ocultar thumbs Ver también para 758 KFD Titrino:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

3.3.2 Pin Assignment

RS232C Interface
Transmitted Data (TxD).
If no data are transmitted, the line is held in the "ON" condition. Data
will only be sent when CTS is in the "ON" condition.
Received Data (RxD)
Request to Send (RTS)
ON condition: Titrino is ready to send data.
Clear to Send (CTS)
ON condition: Remote station is ready to receive data.
Signal ground (GND)
Data Terminal Ready (DTR)
ON condition: Instrument is ready to receive data.
758 KFD Titrino
Protective earthing
Direct connection from cable plug to the protective ground of the instrument.
Polarity allocation of the signals
- Data lines (TxD, RxD)
voltage negative (<-3 V): signal state "ON"
voltage positive (>+3 V): signal state "ZERO"
- control or message lines (CTS, RTS, DTR)
voltage negative (<-3 V): OFF state
voltage positive (>+3 V): ON state
In the transitional range from +3 V to -3 V the signal state is undefined.
Driver 14C88
according to EIA RS 232C specification
Receiver 14C89
3.3 Properties of the RS 232 Interface
"
"
Pin 3
Transmitted Data
Pin 2
Received Data
Pin 7
Request To Send
Pin 8
Clear To Send
Pin 5
Signal Ground
Pin 4
Data Terminal Ready