to position the electrodes on the carotid sinuses (carotid) or genitals, in
particular in patient with a well-known sensibility on reflection of the carotid
sinuses; to position the electrodes near genitals and in those areas that have
poor sensibility;
•
to stimulate the thyroid or apply stimulation on the neck and mouth, as this
stimulation could cause important muscle spasms that can obstruct the
airways, creating difficulty in breathing and problems with the heart rhythm
and blood pressure;
•
to use pointed or sharp objects on the device keyboard.
Warning:
•
insufficiently sized electrode sections can cause skin reactions or burns. Only
use the round electrodes supplied by the manufacturer for facial
applications;
•
do not use damaged electrodes even if they well adhere to the skin;
•
be sure that the electrodes well adhere to the skin. Repeated use of the same
electrodes can compromise the safety of the stimulation, in fact it can cause
skin redness that can last for many hours after stimulation;
•
pay attention to use connection cables with children/young people:
strangulation danger;
•
do not mix connection cables up with earphones or other devices and do not
connect the cable to other equipment;
•
keep right distance between electrodes: the contact between electrodes
could cause wrong stimulations or irritations/burns;
•
stimulation intensity and electrodes position should be suggested by the
prescriber doctor;
•
during treatment with round and 41x41 mm electrodes, current density may
be in excess of 2 mA/cm² for each electrode. In these cases, take special care
regarding any skin redness.
The manufacturer considers himself responsible for the performances, reliability,
safety and security of the device only if:
•
any addition, modification and/or repair are carried out by authorized
personnel;
•
the environmental electrical installation to which MIO-CARE is connected
is compliant to the national laws;
•
the instructions for use contained in this manual are strictly followed.
Should any foreign materials penetrate the device contact the retailer or
manufacturer immediately. If dropped down, check that the housing is not cracked
or damaged in any way; if so, contact the retailer or manufacturer.
IACER Srl
71
MNPG95