Info text
(is displayed in case of
Error
fault when the Info key is
pressed)
ERR 99
DEVICE FAILURE
ERR 98
CONFIGURATION ERROR
ERR 97
NO MODULE INSTALLED
ERR 96
WRONG MODULE
ERR 95
SYSTEM ERROR
ERR 01
NO SENSOR
ERR 02
WRONG SENSOR
ERR 03
CANCELED SENSOR
Error codes
Error Messages
Problem
Possible causes
Error in factory settings
EEPROM or RAM defective
This error message only occurs
in the case of a total defect. The
device must be repaired and
recalibrated at the factory.
Error in confi guration or
calibration data
Memory error in device program
Configuration or calibration data
defective; completely reconfig-
ure and recalibrate the device.
No module
Please have the module installed
in the factory.
Wrong module
Please have the module replaced
in the factory.
System error
Restart required.
If error still persists, send in the
device for repair.
pH sensor *
Sensor defective
Sensor not connected
Break in sensor cable
Wrong sensor *
Sensor devaluated *
17