Technical Information
Interconnections
Pin 3-8 must be shorted to allow the Turbo-V 1000HT pump to start if no
NOTE
interlock contact is connected. If, after starting the pump, the interlock contact
opens, it has no effect on the operation and the pump continues to turn.
The following figure shows a typical contact logic input connection
and the related simplified circuit of the controller.
Figure 9

Connection B J2 - Logic Output Interconnections

Figure 10
212/262
Typical logic input connection
J2 OUT
15
FAULT (+24V)
R 2 (+24V)
ST ART (+24V)
LOW SPEED (+24V)
R 1 (+24V)
SPEED MONITOR (0-10 V)
9
Logic output connector
Turbo-V 1000HT PCB Controller User Manual / 87-900-906-01
8
FAULT
R 2
ST ART
LOW SPEED
R 1
POWER MONITOR (1 Vdc = 50 W)
POWER AND SPEED MONITOR (COM)
1
(F)