Normalized Symbols
This device is a FDA Class II device unit with its own internal electrical
power, with type BF applied sections. IMPORTANT: The information
in this manual must be strictly observed. The ON/OFF switch is a multi-
function key.
Operating Conditions
•
Temperature:
•
Relative Humidity:
•
Atmospheric Pressure:
Storage and Transport
•
Temperature:
•
Relative Humidity:
•
Atmospheric Pressure:
The BioniCare is rated IP22 in accordance with IEC60529. This rating
means that the device is protected against ingress of foreign objects
larger than 12.5mm and against ingress of water with harmful affects when
exposed to dripping at 15 degree tilt.
5° C - 40° C
15% - 93%
700 to 1060 hPa (0.69 to 1.05 atm)
-25°C - +70°C
30% - 93%
700 to 1060 hPa (0.69 to 1.05 atm)
ENGLISH | 19