BioniCare
EMC Compliance Table
Medical electrical equipment needs special precautions regarding EMC and needs to
be installed and put into service according to the EMC information provided below.
Portable and mobile RF communications equipment can affect medical electrical
equipment.
Use of lead wires and other accessories other than those supplied by VQ OrthoCare
may result in increased emissions or decreased immunity of the BioniCare Hand
System.
Guidance and manufacturer's declaration - electromagnetic emissions
The BioniCare Hand System is intended for use in the electromagnetic environment
specified below. The customer or user of the BioniCare Hand Systems should assure that it is
used in such an environment.
Emissions test
RF Emissions
RF Emissions
Harmonic emissions IEC
61000-3-2
Voltage fluctuations /
flicker emissions IEC
61000-3-3
Guidance and manufacturer's declaration - electromagnetic immunity
The BioniCare Hand System is intended for use in the electromagnetic environment
specified below. The customer or user of the BioniCare Hand System should assure that
it is used in such an environment.
Immunity test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient / burst
IEC 61000-4-4
20
®
Hand System
Compliance
The BioniCare Hand System uses RF energy only for
its internal function. Therefore it RF emissions are
Group1
very low and are not likely to cause any interference
Class B
Complies
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Complies
IEC 60601
Test Level
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
± 1 kV for input /
output lines
Electromagnetic Environment - Guidance
in nearby electronic equipment.
The BioniCare Hand System is suitable for
use in all establishments, including domestic
Compliance level
± 6 kV contact
± 8 kV air
Not Applicable
Electromagnetic
environment -
guidance
Floors should be
wood, concrete or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.