SYSTEM ERROR
The maximum warm-up time has
been exceeded.
Please check the sensor block and
the ambient temperature of the
analyzer assembly.
SYSTEM ERROR
Zero calibration is not possible.
Please check the zero gas path in
the analyzer assembly.
SYSTEM ERROR
Span calibration is not possible.
Please check the span gas path in
the analyzer assembly.
SERVICE REQUIREMENT
The service interval has expired.
The system must be serviced.
SYSTEM ERROR
The measured value is not
plausible.
SERVICE REQUIREMENT
Zero drift is out of tolerance.
Please check the sample gas filter
and gas-carrying parts in the
analyzer assembly.
SERVICE REQUIREMENT
Span drift is out of tolerance
Please check the sample gas
capillary and gas-carrying parts in
the analyzer assembly.
SERVICE REQUIREMENT
The temperature of the sensor
block is instable.
Please check the heating of the
sensor block.
SYSTEM ERROR
Zero drift is out of tolerance.
Please check the sample gas filter
and gas-carrying parts in the
analyzer assembly.
SYSTEM ERROR
Span drift is out of tolerance
Please check the sample gas
capillary and gas-carrying parts in
the analyzer assembly.
SYSTEM ERROR
The temperature of the sensor
block is out of tolerance.
Please check the sensor block
temperature and the ambient
temperature.
EuroFID3010 Inline UEG · Instrucciones de servicio · 8017004/ZXA4/V2-2/2018-04 · © SICK AG
▸
Notificar el Servicio al cliente
Fallo
de SICK.
▸
Notificar el Servicio al cliente
Fallo
de SICK.
▸
Notificar el Servicio al cliente
Fallo
de SICK.
▸
Notificar el Servicio al cliente
Petición de
de SICK y tomar las medidas
mantenimiento
de mantenimiento
preventivas.
▸
Notificar el Servicio al cliente
Fallo
de SICK.
▸
Petición de
Notificar el Servicio al cliente
mantenimiento
de SICK.
▸
Comprobar la concentración
Petición de
del gas de prueba en el
mantenimiento
cilindro de gas.
▸
Petición de man-
Notificar el Servicio al cliente
tenimiento
de SICK.
▸
Notificar el Servicio al cliente
Fallo
de SICK.
▸
Comprobar la concentración
del gas de prueba en el
cilindro de gas.
Fallo
▸
?Si esto no ha servido de
ayuda: notificar el Servicio al
cliente de SICK.
▸
Comprobar la temperatura
ambiente en el analizador.
▸
Fallo
Si esto no ha servido de
ayuda: notificar el Servicio al
cliente de SICK.
Solución de problemas