J.P. SELECTA Autester ST Dry PV III Manual De Instrucciones página 27

Autoclave con ciclos de vacío
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES
WATER
ACTIVATE
SETUP > PROBE
LANGUAGE
TIME/DATE
PRINTER
UNITS
PROBE
WATER
SETUP
Access to calibration and
configuration parameters.
LANGUAGE
TIME/DATE
PRINTER
UNITS
PROBE
WATER
SETUP
PROGRAMACIÓN
AJUSTES
PROGRAMMING
SETTINGS
TRABAJO
TASK
Cycle
number
PROGRAMM O
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
J
.P. SELECTA s.a.u.
CODIGO 80330 REV B
7.5.6
With this parameter set to «activate» the autoclave will take water from an
external network for the steam generation.
With this parameter set to «disabled», the autoclave will take water from the
tank itself, for the steam generation.
Press
BACK
parameter.
HH:MM:SS DD/MM/YY
7.5.7
PASSWORD
The password protects the parameters that may affect the operation of the auto-
clave. It is only accessible by contacting to J.P. Selecta, s.a.u. technical service.
CHANGE
PASSWORD
There are two groups of parameters protected by a password:
CAL
TEST
BACK
HH:MM:SS DD/MM/YY
7.6
The autoclave has the EN13060 test cycles to made easy operation qualification:
PASSWORD
vacuum according to EN13060.
CHANGE
PASSWORD
into wrapped loads.
This test cycles should be run periodically, and results must be stored as quality
system records.
CAL
TEST
BACK
HH:MM:SS DD/MM/YY
Test
cycles
7.7
Up to 8 cycles could be stored on the autoclave.
The cycle sets the treatement to apply over samples, by means of the setting
parameters.
cycle
selection
7.7.1
icons
- Select the cycle to modify by pointing over:
- The cycle (program) sets the treatement to apply over samples, by means of
SEL
the setting parameters.
C
O
SAVE
HH:MM:SS DD/MM/YY
Note: Pointing over this icon
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Septiembre 2017
(Sujetas a modificaciones sin previo aviso) Pag.: 27
Activates /disables the water entry
to return to themain screen or
Parameters protected by a password
Calibration of meassuring devices:
PROBE: Offset value for heart temperature probe.
WATER: Offset value for autoclave water temperature probe.
Offset value for pressure sensor.
Temperature for strip heaters.
Set of parameters which configurates the autoclave operation.
Test cycles (Vacumm test & Bowie Dick)
Vacumm test: Qualifies the autoclave tightness and its capabilty to get suitable
BOWIE DICK test: Qualifies the steam quality and steam capability to penetrate
Sterilization cycles (Programs)
Setting cycles:
-
Purge (air removal) type.
-
Sterilization temperature.
-
Sterilization time.
-
Drying time.
BACK
will use selected cycle as a default.
SEL
Tel 34 93 770 08 77 Fax 34 93 770 23 62
to modify another

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4002420

Tabla de contenido