J.P. SELECTA Autester ST Dry PV III Manual De Instrucciones página 28

Autoclave con ciclos de vacío
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL
PROGRAMA O
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
Purge type
PROGRAMA O
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
Sterilization temperature
PROGRAMA O
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
Sterilization time
PROGRAMA O
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
Drying time
PROGRAMA O
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
Save program changes
PROGRAMA O
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
SETUP
Quit from main menu
J
.P. SELECTA s.a.u.
CODE 80330 REV B
September 2017 (It can be modified without notice)
SEL
Set the air removal type:
C
O
Solid 1: 1 vacuum cycle.
Solid 2: 2 vacuum cycles. For wrapped loads.
Solid 3: 3 vacuum cycles. For double wrapped loads,
SAVE
Liquid
HH:MM:SS DD/MM/YY
SEL
C
O
Press over temperature label to set open the range from 115ºC ... 134ºC
SAVE
HH:MM:SS DD/MM/YY
SEL
Press over sterilization time to set time from 1 to 59 minutes.
C
O
SAVE
HH:MM:SS DD/MM/YY
SEL
Press over sterilization time to set time from 1 to 59 minutes.
C
O
SAVE
HH:MM:SS DD/MM/YY
SEL
C
O
SAVE
HH:MM:SS DD/MM/YY
SEL
C
O
SAVE
HH:MM:SS DD/MM/YY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Setting air removal operation (purge).
For not wrapped loads.
catheters, hollow devices, ...
Not vacuum.
Culture media in demi-close bottles.
Setting sterilization temperature.
Setting sterilization time.
Setting drying time.
Store programs.
Return to main menu.
Page: 28