Operation; Select Program - J.P. SELECTA Autester ST Dry PV III Manual De Instrucciones

Autoclave con ciclos de vacío
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

MANUAL DE INSTRUCCIONES
PROGRAMA 0
PRO.: SÓL. 2
TEMP.: 134
T.EST.: 4m
T.SEC.: 10m
ACTUAL: +0.00 bar 24ºC
Tiempo de secado.
Shows running
programm status
PROGRAMM 0
-0.70 bar
24 ºC
TIME: 00:00:06
S2 - 134ºC - EST 4' SEC 10'
STERILIZATION
CORRECT
DOOR OPEN IS ENABLED
ACTUAL: +0.00BAR 23ºC
J
.P. SELECTA s.a.u.
CODIGO 80330 REV B

7. Operation

7.1 Load
Check tank water level.
Load the previously cleaned and settled material baskets.
Check that the seal is in place.
If it is installed and if it is necessary, place the heart probe in the choosed refe-
rence point.
Close the lid.

7.2 Select Program

Select the suitable program for sterilization and drying the load.
To set the program (cycle) see 7.3.
C
O
Check the USB device registration and configuration.
Start the cycle.
RUN
HH:MM:SS DD/MM/YY
7.2.1 Application of the cycle
During the cycle (also called software) autoclave screen will indicate the segment
that is being made.
Cycle Segments are:
STEP: WATER
WATER:
VACUUM:
HEAT
STERILE.
EMPTY:
STOP
UNPRESS :
HH:MM:SS DD/MM/YY
OK:
FAIL:
OK
This is the "End of cycle" message.
HH:MM:SS DD/MM/YY
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Septiembre 2017
(Sujetas a modificaciones sin previo aviso) Pag.: 23
The autoclave lid only opens with the autoclave
connected to the mains ( Main Switch lighted )
Do not force the opening lever
Water inlet to the boiler.
Removing the air.
Steam generation.
Sterilization of the load.
Draining the boiler steam and water.
Depressurization boiler.
Sterilization conditions have been achieved according to
cycle parameters.
Not achieved all the conditions for the sterilization.
Tel 34 93 770 08 77 Fax 34 93 770 23 62