Maintenance - J.P. SELECTA Autester ST Dry PV III Manual De Instrucciones

Autoclave con ciclos de vacío
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

INSTRUCTION MANUAL
J
.P. SELECTA s.a.u.
CODE 80330 REV B
September 2017 (It can be modified without notice)

11. Maintenance

Before removing the lid, disconnect the machine from the
mains.
The control board should only be changed by a technical ser-
vice or authorized personnel.
CLEANING:
ducts.
alcohol.
Monthly operations:
Three months operations:
Yearly operations:
temperature standard.
Autovía A-2 Km 585.1 Abrera 08630 (Barcelona) España
e-mail: selecta@jpselecta.es - website: http://www.jpselecta.es
Wall panels and stainless steel: Use steel cleaning home pro-
Limescale deposits: Use home cleaning products.
TFT Screen: use glasses cleaning products.
Cover and plastic areas: use non-abrasive cloth dipped in
Clean front bottom drainage filter.
Change re-used water from water reservoir.
Record results from vacuum test and Bowie dick test.
Change air entry filter (Autoclave rear side)
Change re-used water from water reservoir.
Check and contrast temperature autoclave against traceable
Check, visually the state of door silicone gasked.
Check safety valve operation.
ATTENTION: this operation should be run for qualified
personnel.
Dangerous contact with hot steam.
Check, visually the state of electrical wiring, ground and mains.
Page: 32
Tel 34 93 770 08 77 Fax 34 93 770 23 62

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

4002420

Tabla de contenido