10. Safety Systems
10.1 Safety pressure switch
When an abnormal overpressure is produced in the sterilization cham-
ber, the safety pressure switch acts in a way to stop the autoclave and
it opens the aireation electrovalve.
10.2 Safety valve
If an abnormal overpressure is produced, the safety valve is opened
(10), expelling the steam.
This valve is calibrated to initiate the opening between 2.2 kg/cm²
and 2.5 kg/cm².
In case of safety valve reset, the machine must be prepared to advise
a technical personnel.
10.3 Safety thermostat
When an important increase or decrease of water level is produced
the safety thermostat (14) acts by stopping the heater.
In case of open lid blocked by a safety system failure make as follows:
1 Stop the machine.
2 Disconnect the machine from the mains.
3 Place an outlet and a tube in the aireation valve (11).
4 Carefully open the valve to avoid steam leackages. Be careful to
avoid possible burns.
5 Wait for the total steam emptying.