Hach Modbus Manual Del Usuario página 12

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 29
Table 7 Classified status 3 – register 49933 (continued)
Bit
Error
3
4
Measurement 8 Quality Bad
5
Measurement 8 Low Limit
6
Measurement 8 High Limit
7
Measurement 9 Quality Bad
8
Measurement 9 Low Limit
9
Measurement 9 High Limit
10
Measurement 10 Quality
Bad
11
Measurement 10 Low Limit
12
Measurement 10 High Limit
13
Measurement 11 Quality
Bad
14
Measurement 11 Low Limit
15
Measurement 11 High Limit
Table 8 Classified status 4 – register 49934
Bit
Error
0
Reserved for future use
1
2
3
12 English
Note
Precision of measurement is out of
specified limits.
Measurement is below measurement
range.
Measurement is above measurement
range.
Note
Fixed at 0
Table 8 Classified status 4 – register 49934 (continued)
Bit
Error
4
Measurement 12 Quality
Bad
5
Measurement 12 Low Limit
6
Measurement 12 High Limit
7
Measurement 13 Quality
Bad
8
Measurement 13 Low Limit
9
Measurement 13 High Limit
10
Measurement 14 Quality
Bad
11
Measurement 14 Low Limit
12
Measurement 14 High Limit
13
Measurement 15 Quality
Bad
14
Measurement 15 Low Limit
15
Measurement 15 High Limit
Note
Precision of measurement is out of
specified limits.
Measurement is below measurement
range.
Measurement is above measurement
range.

Publicidad