Using a #1 Phillips Head Screw Driver and furnished screws connect the power cord retention bracket
to the IntraTemp Solution Warmer via the threaded connectors located to each side of the power inlet
module.
8.4

IntraTemp Basin Temperature Verification:

The IntraTemp basin solution temperature should be verified every five (5) years. The following items
are required to perform this verification:
-
Calibrated Fluke 51II Digital Thermometer or equivalent.
-
Shrouded 6" long Type T Thermocouple Probe.
Install a ThermaBasin. Add two (2) liters of saline or water into the ThermaBasin. Turn the unit on and
allow the fluid to heat for two (2) hours.
Place the tip of the Thermocouple Probe into the fluid (without contacting the basin) at a point ~1" (2.5 cm)
above the center of the basin. The temperature indicated by the calibrated thermometer should be at Set-
Point ±1°C (±1.5°F). If the temperature is not within the specified range, contact Ecolab/Microtek
Customer Service.

9. CONTACT INFORMATION

Ecolab / Microtek Customer Service
North America: ......(800) 824-3027
Europe: ................ +31 575 599200
Manufactured for:
Microtek Medical, Inc.
602 Lehmberg Road
Columbus, MS 39702 U.S.A.
www.microtekmed.com
Microtek Medical B.V.
Hekkehorst 24
7207 BN Zutphen, The Netherlands
©2014 ECOLAB INC. or its affiliates All rights reserved
13 of 128
Figure 9: Fuse Drawer Opening

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Itw6l

Tabla de contenido