Heater Safety Switch; Temperature Control System; Upper Basin Indicator Light And Digital Display; Heated Bottle Pocket Technology - Ecolab IntraTemp Manual Del Operador

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 37
3.4

Heater Safety Switch

The ThermaBasin heating plate Safety Switch Thermocouple has three positions:
Fully up- no ThermaBasin in place - no heat
Middle position- ThermaBasin in place - heat conducted to ThermaBasin
Fully down - no heat
Care should be exercised that the ThermaBasin is properly located in the basin receptacle with the basin's
thermal well placed securely over the warmer's Safety Switch thermocouple.
3.5

Temperature Control System

To ensure the safe use of warm fluids the basin heater controls may be set at a temperature between
37° and 43°C (98° - 110°F) and the temperature scale may be set to either Celsius or Fahrenheit by
qualified service personnel. Access to the control buttons by the user is restricted by a controller cover.
3.6

Upper Basin Indicator Light and Digital Display

The green "Power On" indicator light glows to indicate that power is available to the upper basin heater and
that a ThermaBasin is properly installed for use.
The digital display on the top surface of the IntraTemp displays the temperature of the solution in the
fluid basin.
If the ThermaBasin is not properly seated, neither the green indicator light nor the digital display will be
activated; the basin heating system is not actuated.
3.7

Heated Bottle Pocket Technology

The side housing of the solution warmer has two (2) Bottle Pocket heated access sleeves to hold standard 1
or 1.5 liter sterile saline solution bottles or 1 liter bags.
The Bottle Pockets cannot control a fluid bottle or bag to an exact temperature; they are designed
so that when a fluid container is in place it will be maintained at or very near a steady temperature.
3.8

Drain Tank

A Drain Tank is provided to capture any fluid spilled into the upper housing receptacle. A tube that runs
inside the unit connects the drain hole located in the upper housing receptacle to the drain tank.

4. UNPACKING INSTRUCTIONS

4.1

Unpacking the IntraTemp Solution Warmer:

1. DO NOT CUT THE BOX. Cut the strapping bands and remove.
2. Cut the tape holding the top of the box closed. Lift the box flaps.
3. Remove the Drain Tank from the top of the packaging material and set aside for installation onto the unit.
4. Remove the packaging material from the top of the unit.
5. Lift the box off the IntraTemp machine.
6. Remove IntraTemp from the lower packaging.
7. Save the shipping material if a unit is to be returned.
NOTE: Locate the Power Cord Kit supplied in a separate box.

5. INSTALLATION INSTRUCTIONS

NOTE: Procedures described in this section should only be performed by qualified service personnel or Ecolab/
Microtek Factory Representative
5.1

Tools Required:

The procedures described in this section will require the following tools not included in the IntraTemp Solution
Warmer packaging:
-
#1 Phillips Head Screw Driver
-
#2 Phillips Head Screw Driver
8 of 128

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Itw6l

Tabla de contenido