9. Product Specifications
9.1 Delivery form
– Programat CS
– Power cord
– Vacuum hose
– Calibration Test Pack
– Operating Instructions
– Programat Firing Tray Kit
– Firing tongs
Recommended accessories
– Programat Accessories Set
– Automatic Temperature Checking Set 2 (ATK 2)
– Vacuum Pump VP3 easy
9.2 Technical data
Power supply
Overvoltage category II
Contamination level 2
Tolerated voltage fluctuations
Max. power consumption
Acceptable data for vacuum pump of other manufacturers
Max. output:
250 W / max. leakage current 0.75 mA
Final vacuum:
Electrical fuses:
Dimensions of electrical fuses
Dimensions of the closed furnace
Depth: 430 mm / width: 305 mm / 410 mm (with Cooling tray)
Usable size of the firing chamber
Max. firing temperature
Weight
Safety information
The CS complies with the following guidelines:
– IEC 1010-1/EN 61010, Part 1
– UL and cUL standards
Radio protection / electromagnetic compatibility
100 V / 50–60 Hz
110–120 V / 50–60 Hz
200–240 V / 50–60 Hz
+/- 10%
11 A at 100 V
12 A at 110–120 V
8.5 A at 200–240 V
< 50 mbar
Use only tested pumps
100 V / 110–120 V:
250 V / T 15 A (heating circuit)
250 V / T 5 A (vacuum pump)
200–240 V:
250 V / T 8 A ( heating circuit)
250 V / T 3.15 A ( vacuum pump)
110–120 V:
Diameter 6.3 x 32 mm
200–240 V:
Diameter 5 x 20 mm
Height: 320 mm
Diameter 80 mm
Height 48 mm
1200 °C (2192 °F)
Furnace base: 12.0 kg
Furnace head: 4.5 kg
EMC tested
9.3 Acceptable operating conditions
Acceptable ambient temperature range:
+5 °C to +40 °C (+41 °F to +104 °F)
Acceptable humidity range:
80 % maximum relative humidity for temperatures up to 31 °C
(87.8 ° F) gradually decreasing to 50 % relative humidity at 40 °C
(104 °F); condensation excluded.
Acceptable ambient pressure:
The furnace is tested for use at altitudes of up to 2000 m above sea
level.
9.4 Acceptable transportation and storage conditions
Acceptable temperature range
Acceptable humidity range
Acceptable ambient pressure
Use only original packaging of the Programat CS together with the
respective foam material for shipping purposes.
30
-20 to +65 °C (-4 °F to +149 °F)
Max. 80 % relative humidity
500 mbar to 1060 mbar