MSA ALTAIR 5 Manual De Funcionamiento página 37

Ocultar thumbs Ver también para ALTAIR 5:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

The Calibration Options menu allows the user to:
modify the calibration cylinder settings
(CYLINDER SETUP)
enable/disable calibration due date and set the number of days
(CAL DUE OPTIONS)
enable/disable the option to show the last cal date at turn-ON
(LAST CAL DATE)
enable/disable the option for password-protected calibration
(CAL PASSWORD).
1. Press the:
• ▼ button go to next page
• ▲ button to go previous page
button to enter setup.
Setting Calibration Cylinder
This option has a dialog similar to the span calibration dialog.
The display shows all active sensors.
1. Press the
button to enter setup.
• The screen for the first calibration cylinder displays.
2. Press the:
• ▼ or ▲ button to change the value.
button to confirm the setup.
With this confirmation, the instrument automatically moves to the
next cylinder setting.
3. Repeat the sequence for changing the required settings for all
necessary gas values.
• After the last setting is performed, the instrument returns to the
Calibration Options menu.
3-7

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Altair 5 ir

Tabla de contenido