2.4.4 Time Weighted Average (TWA Page)

If the TWA alarm activates, leave the contaminated area
immediately; the ambient gas concentration has reached the
preset TWA alarm level. Failure to follow this warning will
cause over-exposure to toxic gases and persons relying on
this product for their safety could sustain serious personal
injury or death.
This page shows the average exposure over 8 hours since the
instrument was turned ON or the TWA reading was reset. When the
amount of gas detected by the instrument is greater than the TWA limit:
MONOCHROME DISPLAY
TWA
Alarm sounds
Alarm lights flash
"TWA ALARM" message flashes.
To Reset the TWA Readings
1. Access the TWA Page.
2. Press the
button.
The TWA alarm is calculated over an eight-hour exposure.
TWA Calculation Examples:
1-hour exposure of 50 ppm:
(1 hour x 50 PPM) + (7 hours x 0 PPM)
4-hour exposure of 50 ppm and
4-hour exposure of 100 ppm:
(4 hours x 50 PPM) + (4 hours x 100 PPM)
12-hour exposure of 100 ppm:
(12 hours x 100 PPM)
8 hours
This page can be de-activated through MSA Link.
2-12
" WARNING
COLOR DISPLAY
8 hours
8 hours
= 150 PPM
= 6.25 PPM
= 75 PPM