Cyanide
(158)
10 ml
24 mm
Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
1) Fill a cleaned cell (24mm) with 8ml deionised water.
2) Add precisely 2ml sample water to the same cell.
3) Replace the lid on the cell and swivel this back and forth 5 times.
4) Do you want to use the last stored ZERO value?
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
5) Close the lid, place the cell into the photometer (use caution to the
arrow on the cell) and press ZERO.
4
6) Remove the cell and open the lid.
7) Add 2 scoops of PL Cyanide-11.
8) Replace the lid on the cell and swivel this back and forth 5 times.
9) Add 2 scoops of PL Cyanide-12.
10) Replace the lid on the cell and swivel this back and forth 5 times.
11) Add 3 drops PL Cyanide-13 to your sample. Shake the cell to mix the
liquid.
12) Close the lid. Gently shake the cell to mix the liquid. Press TEST.
4
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
78
0.01 - 0.50 mg/l
Name on device: 158-Cyanide-pow
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
PL Cyanide-11
PL Cyanide-12
PL Cyanide-13
(all included in PL158-kit)
= continue with step 5)
= continue with step 7)
Continued...