Iron (MR)
(29)
Continued...
8) Add 1 x 0.05mL (scoop) "PL Iron MR 1" powder to the sample water in
the cuvette.
9) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
10) After the lapse of a 3 minute countdown the determined result is
displayed. Press button 4 to save the test result together with date and
4
time on the device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
B) Measurement procedure for "total iron"
1) Fill 10ml unfiltered sample water into a clean 24mm cuvette.
One-Time-Zero:
Measure new ZERO
Use last ZERO
1
2
3
!
2) Replace the screw lid on the cuvette and insert it into the PrimeLab
sampling chamber. Align the arrow on the front of the cuvette. Press
4
ZERO.
3) Remove the cuvette from the chamber and unscrew the lid.
4) Add 1 x 0.05mL (scoop) "PL Iron MR 1" powder to the sample water in
the cuvette.
5) Replace the lid on the cuvette, swirl it carefully and insert it into the
sampling chamber of the PrimeLab. Align the arrow on the front of the
4
cuvette. Press TEST.
6) After the lapse of a 3 minute countdown the determined result is
displayed. Press button 4 to save the test result together with date and
4
time on the device.
125
0 - 10 mg/l
Name on device: 29-Iron-MR-pow
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
If after switching on the device you have already performed the
ZERO (baseline) adjustment for the same type, the following
message is displayed:
-> Measure new ZERO
-> Use last ZERO
4
For the sake of precision select "Use last ZERO" only
if the stored ZERO value is used with the
same cuvette and the same water sample for the test!
2+/3+
(Fe
)
= continue with step 2)
= continue with step 4)
Continued...
Powder
125