Descargar Imprimir esta página

PrimeLab 1.0 Multitest Manual De Usuario página 68

Desktop assistant
Ocultar thumbs Ver también para 1.0 Multitest:

Publicidad

Idiomas disponibles

Idiomas disponibles

Chloride
(10)
Continued...
10) Press button 4 to save the result together with date and time in the
4
device.
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
(*) The test result can be converted to the following units:
NaCl
->
Avoid severe shaking of the water sample after adding the reagent, as this
can lead to incorrect measurements.
->
The reagent used will cause fine clouding.
->
Other substances in the water that may react with silver nitrate in
an acidic medium will lead to a falsification of the measurement result.
Such species are bromide and iodine.
->
Very alkaline water should be neutralized before the measurement by adding
nitric acid.
35
0.5 - 25 mg/l
Name on device: 10-Chloride-tab
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
-
(Cl )
Tablet
35

Publicidad

Capítulos

loading