Measurement procedure:
The steps up to the selection of the parameter value to be determined are the same for all
procedures and described on page "TEST-5". Please observe the important notes for
accurate measurements on pages "TEST-8" and "TEST-9".
!
The individual steps of the measurement procedure are shown on the display.
1) Rinse vial 3 times with your water sample.
2) Fill a cleaned cell (24mm) with 10ml of your sample.
3) Place the cell into the PrimeLab (use caution to the arrow on the cell).
Do NOT close the lid.
4) Place PTSA-ADAPTER on top of the open cuvette.
5) Press TEST.
4
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Note(s)
->
Ensure that all parts are clean, dry and free of grease and the adapter
must be placed firmly until it stops.
1. As this Parameter uses an indirect light from above, the shape of the bottom of
the cell is important to the result. Because the bottom of the cells can vary
greatly it is imperative to always use the cell with which this parameter was also
calibrated. It is essential to always ensure the correct amount of water in the cell,
which is why exactly 10 ml of liquid should be taken by the pipette for the
subsequent sample measurement. Please Change or clean the tip of the pipette
after each measurement/calibration.
2. Calibrate this parameter via the calibration SET if you use another cell or you
have the feel that the measurement result is inaccurate.
3. One of the following reasons can lead into receiving an error message: "check
adapter"
• Weak or empty batteries (please change)
• Dirty lense (adapter)
• Wrong adapter used for this measurement (there are different adapter for
different measurements, all looking the same)
• Adapter might not stay straight on PrimeLab
• Cuvette-hole (PrimeLab) might be dirty (check the two windows)
• Water sample might be too dark / not enough light can pass water sample
to reach the sensor
270
0 - 1000 µg/l
Name on device: 156-Watch-Ad
PLSp-PIP10ml
Use button 3 (arrow down) to scroll through the steps.
Use button 4 to skip the notes.
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
(Watch)
PLSp-PIP10ml-tips