Phenol
(98)
Continued...
!
Assessment of the measurement value relative to the ideal range defined by you.
Measurement value for this parameter is outside the value range of this method.
Notes
->
If your sample does contain copper, zinc, iron or manganese ions (up to 350
mg/l) add one Phenol CR tablet after ZERO. Crush and mix to dissolve.
->
Level of > 20 mg/l hydrogen peroxide interferere with this test and can
influence the reading.
->
High (free) chlorine levels (> 10 mg/l) interfere with this test and can
influence the reading.
->
Alkalinity above 150 mg/l CaCO as well as sulphite above 10 mg/l or more
than 2 mg/l sulphide will interfere with this test and can influence the reading.
->
Some organic keto-enol compounds can lead to high readings.
188
0 - 5 mg/l
Name on device: 98-Phenol-tab
Possible messages in the line below the measurement value:
"low! / high! / good!":
"Overrange! / underrange!":
3
(C H OH)
6
5
Tablet