PrimeLab 1.0 Multitest Manual De Usuario página 125

Desktop assistant
Ocultar thumbs Ver también para 1.0 Multitest:

Publicidad

Idiomas disponibles

Idiomas disponibles

Fluorescein
(113)
Continued...
Notes
->
(*) Conversion to C H O
->
Ensure that all parts are clean, dry and free of grease and the adapter
must be placed firmly until it stops.
->
As this Parameter uses an indirect light from above, the shape of the bottom
of the cell is important to the result. Because the bottom of the cells can vary
greatly it is imperative to always use the cell with which this parameter was
also calibrated. It is essential to always ensure the correct amount of water
in the cell, which is why exactly 10 ml of liquid should be taken by the
pipette for the subsequent sample measurement. Please change or clean
the tip of the pipette after each measurement/calibration.
->
Calibrate this parameter via the calibration SET if you use another cell or
you have the feel that the measurement result is inaccurate.
->
One of the following reasons can lead into receiving an error message:
"check adapter"
• Weak or empty batteries (please change)
• Dirty lense (adapter)
• Wrong adapter used for this measurement (there are different adapter for
different measurements, all looking the same)
• Adapter might not stay straight on PrimeLab
• Cuvette-hole (PrimeLab) might be dirty (check the two windows)
• Water sample might be too dark / not enough light can pass water sample
to reach the sensor
92
0 - 500 µg/l
Name on device: 113-Fluorescein-Ad
20
12
5
(C H Na O )
20
10
2
5
Adapter
92