Data management

Select a report

The instrument saves measurement, calibration, system health analyses, system suitability tests,
error logs, background statistics and audit record data reports. Use the steps that follow to read or
export a report.
1. Push Reports and Data.
2. Select a report from the list, then push
Option
Description
Filter Results
Selects the data range to sort by date and/or by operator.
Select a format Selects the report template output format (PDF, CSV or XML).
Output
Selects the output device: printer, export to USB, paperless export or show on the display.
To get the report via printer, make sure a printer is connected to the instrument. Refer to
Configure the peripheral settings
To output a report via a USB port, connect a USB flash drive. When prompted, push Start.

Do a qualifications procedure

Chemical exposure hazard. Obey laboratory safety procedures and wear all of the personal protective
equipment appropriate to the chemicals that are handled. Refer to the current safety data sheets
(MSDS/SDS) for safety protocols.
Collect the necessary items before a qualification procedure starts. Do these procedures that follow
periodically or as required by regulatory agencies.
1. Push Qualifications.
2. Select an option.
Option
Description
Calibration
Shows the instructions on the display about how to prepare the reagent and the standards.
SST500
Shows the instructions on the display about how to prepare the reagent and the standards.
Background
Shows the instructions on the display about how to prepare the reagent.
SDBS
Shows the instructions on the display about how to prepare the reagent and the standards.
SST8000
Shows the instructions on the display about how to prepare the reagent and the standards.
Accuracy
Shows the instructions on the display about how to prepare the reagent and the standards.
3. Enter the lot identification number (Lot ID) and the expiration date for the standards.
4. Push
to start the procedure.
If necessary, push Pause or Cancel during the procedure.
2
This step does not apply to the background test.
for output options.
on page 16. When prompted, push Print.