Össur Pro-Flex LP Torsion Instrucciones Para El Uso página 7

Ocultar thumbs Ver también para Pro-Flex LP Torsion:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 19
can cause the device to malfunction if not correctly reassembled.
Assembling the Device
See Figure 6.
N.B.: Bumper size depends upon foot size and category.
AFTER DISASSEMBLY / ASSEMBLY
Check the following:
• Are the duckbill valves oriented in the correct direction (Figure 4)?
• Is the membrane fully inserted into the housing?
• Are the inlet and outlet ports tightened?
• Is the tube fully inserted on the inlet port? (Check both at the socket
valve and the membrane housing)
HOW TO CHANGE THE DEVICE MEMBRANE
1. Use the Flex-foot shoehorn (or equivalent) to remove the prosthetic
foot from the foot cover.
2. Disconnect the tube from the device membrane housing.
3. Untighten the screw that fastens the membrane housing to the Pro-
Flex LP Torsion pyramid.
4. Follow Figure 5.
5. Lubricate the membrane before sliding it into the housing with
Iceross Clean&Simple Spray or equivalent.
FINISHING / LINER OPTIONS
Össur recommends that the device is used with an Iceross Seal-In
Precaution: Make sure the Flex-Foot sock does not interfere with the
device as it could compromise the function of the pump.
THE TUBE
The tube is the most sensitive part of the device, and special attention
should be paid to it for potential functional issues that might arise.
Precaution: Ensure that device tubing is not compressed or folded by the
application of cosmetic finishing.
Ensure the tube is NOT:
• Folded (Figure 3).
• Pressed up against sharp objects.
• Punctured.
• Dangling during gait.
• Located laterally, as there is more likelihood of damage.
TROUBLESHOOTING
Check the following:
• Is there leakage in the socket?
• Is the tube punctured?
• Is the membrane punctured or torn?
• Has the membrane been pulled out of the membrane housing?
• Is there dust or grit inside the device?
• Is the tube fully inserted on the inlet port? (Check both at the device
socket valve or membrane housing)
• Are the inlet and outlet ports tightened?
• Is the device valve leaking?
liner.
®
7

Publicidad

Tabla de contenido
loading

Tabla de contenido