7 CALIBRATING YOUR GAUGE (continued)
To enable and disable calibration check:
1
Press Menu/Calibration.
2
Use the
'Calibration Check' and press
'Select'.
3
To disable, press 'Select' again to
un-check the 'Calibration Check'
radio button.
7.8 LOCKING THE CALIBRATION
Using the 'PIN Lock' feature, the calibration settings can be 'locked',
preventing the user from making any changes to the calibration
without first disabling PIN lock.
Users can still test the calibration via Menu/Calibration/Test
Calibration when 'PIN Lock' is enabled, but are unable to validate or
re-calibrate the gauge.
For more information on 'PIN Lock', see Section 8 'PIN Lock' on
page en-18.
7.9 CALIBRATION MEMORIES - PTG8
Up to three calibrations can be saved in the gauge memory. Once
saved, the user can select the calibration memory - without the need
to re-calibrate the gauge.
To save a calibration into memory:
1
Press Menu/Calibration/Cal Memory n, where n = 1, 2 or 3.
2
Use the
'Select'.
3
Use the
method and follow the on-screen instructions to calibrate the
gauge.
4
The calibration will be stored in the gauge memory as Cal
Memory n, where n = 1, 2 or 3.
en-17
R
softkeys to highlight
softkeys to highlight 'Cal Method' then press
softkeys to highlight the required calibration
www.elcometer.com