Vacuson 60LP
Please pay attention to the following important points for the maintenance of the material :
•
To avoid damage, do not pull cable when disconnecting handpiece and
electronicmotor!
•
Do not use dissolving agents for cleaning, in particular agents with ingredients
as Methylene chloride and Trichloroethane!
•
Autoclave material in sterilization bag!
•
Autoclave at maximum 135°C!
•
Sterilized material should be stored and tagged with sterilizing date!
6.1 Control unit and foot control
The control unit and the floor pedal do not come into contact with patients. Only clean the surface of
the unit, do not use harsh cleaners or solvents for cleaning. Use 80% ethyl alcohol or microbiologically
effective disinfectants.
The Control Unit should be wiped with a clean, damp cloth only. No spray cleaners
or disinfecting agents should be employed, as the Control Unit face is not
watertight!
10
6 Cleaning, disinfection, and sterlization